新制和失效钯炭催化剂中钯含量的ICP-AES法测定 |
Determination of Palladium in Fresh and Spent Pd/C Catalyst by ICP-AES |
Received:March 06, 2012 |
DOI: |
中文关键词: 分析化学 钯炭催化剂 ICP-AES 钯 |
英文关键词: analytic chemistry Pd/C catalyst ICP-AES palladium |
基金项目: |
Author Name | Affiliation | E-mail | LIU Wei | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | liuweiynkm@sina.com | CAI Wenyun | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | | MA Yuan | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | | WANG Yingjin | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | | LI Guangli | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | | XU Guang | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | | HE Jiao | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | | FANG Haiyan | State Key Laboratory of Advanced Technology for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China | |
|
Hits: 5025 |
Download times: 5581 |
中文摘要: |
试样采用HNO3-HClO4湿法消解、或在700℃下灼烧除去碳,王水溶解残渣后用ICP-AES测定钯含量。考察了样品前处理、方法精密度和准确度,样品加标回收率为99%~101%,相对标准偏差<1%。方法准确、快速、简便。 |
英文摘要: |
The content of Pd was determined through (1) wet digestion with HNO3-HClO4 or ignition at 700℃ to remove carbon, (2) dissolution of the residues by aqua regia and (3) determination of palladium by ICP-AES. Pretreatment of sample, accuracy and precision of the method were discussed. The additional standard recovery is 99%~101%, RSD <1%. The method is simple, accurate and fast. |
View Full Text
View/Add Comment Download reader |
Close |
|
|
|