Abstract
微波消解–ICP-AES法测定Pd950首饰中多种有害元素
Determination of Several Harmful Elements in Pd950 Jewelry by Microwave Digestion–ICP-AES Method
Received:February 10, 2015  
DOI:
中文关键词: 分析化学  微波消解  电感耦合等离子体-原子发射光谱法  Pd950首饰          
英文关键词: analytical chemistry  microwave digestion  ICP-AES method  Pd950 jewelry  Hg  Pb  Cd  Cr  As
基金项目:山东省质量技术监督局科研项目(2011KY02)。
Author NameAffiliationE-mail
ZHANG Fengxia National Gold and Diamond Testing Center of China, Shandong Institute of Metrology,Shandong Social Justice Institute of Metrology, Ji’ nan 250014, China zhfenxia@163.com 
CHENG Youfa National Gold and Diamond Testing Center of China, Shandong Institute of Metrology,Shandong Social Justice Institute of Metrology, Ji’ nan 250014, China  
WANG Ping National Gold and Diamond Testing Center of China, Shandong Institute of Metrology,Shandong Social Justice Institute of Metrology, Ji’ nan 250014, China  
YAN Fei National Gold and Diamond Testing Center of China, Shandong Institute of Metrology,Shandong Social Justice Institute of Metrology, Ji’ nan 250014, China  
Hits: 4313
Download times: 4259
中文摘要:
      提出了试样经微波消解后用电感耦合等离子体原子发射光谱法(ICP-AES)测定Pd950首饰中汞、铅、镉、铬、砷的分析方法。在120℃温度下,以王水为消解试剂,试样在密闭容器中消解25 min可以完全溶解。汞、铅、镉、铬、砷元素的检出限为0.006~0.15 mg/kg,回收率为92%~116%,RSD为0.07%~1.4%。总分析流程均大大缩短,微波消解前处理方法具有简便、快捷、回收率高、精密度高等优点,能够满足测定Pd950首饰中汞、铅、镉、铬、砷的需要。
英文摘要:
      A method for determination of Hg, Pb, Cd, Cr, As in Pd950 jewelry by inductively coupled plasma atomic emission spectrometry (ICP-AES) after microwave digestion was established. The results showed that the sample could be fully dissolved in aqua regia by microwave digestion for 25 min in closed container at 120℃. The detection limit was 0.006~0.15 mg/kg. The recovery rate was 92%~116% and the relative standard deviations (RSD) of the method was 0.07%~1.4%. The microwave digestion method is simple, rapid with high recovery rates and accuracy, and is applicable for the determination of Hg, Pb, Cd, Cr, As in platinum jewelry.