Abstract
络合滴定法测定化合物中钯量的不确定度评定
Evaluation of Uncertainty for Palladium Content in Palladium Compounds by Complexometric Titration
Received:June 17, 2015  
DOI:
中文关键词: 分析化学  不确定度  络合滴定  钯化合物
英文关键词: analytical chemistry  uncertainty  complexometric titration  palladium compounds
基金项目:
Author NameAffiliationE-mail
XING Yinjuan State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China zhulou@sina.com.cn 
GAN Jianzhuang State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China  
JIN Yaqiu State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China  
ZENG Hefeng State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China  
MA Yuan State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China  
TAO Saixiang State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China  
YANG Meiying State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China  
QIAN Yanlin State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals,Sino-Platinum Metals Co. Ltd., Kunming 650106, China  
Hits: 2904
Download times: 3237
中文摘要:
      对二甲基乙二醛肟析出-EDTA络合滴定法测定系列钯化合物中钯量的不确定度进行了分析与评定,并计算出扩展不确定度。结果表明,测定二氯化钯、醋酸钯、二氯四氨钯、硝酸钯溶液和硫酸钯溶液中59.59%、47.73%、42.46%、17.64%和4.01%的钯含量,扩展不确定度分别为0.30%、0.26%、0.24%、0.10%和0.03%。不确定度主要来源于终点判断和标定过程。
英文摘要:
      The uncertainty for complexometric titration of EDTA to determine the palladium contents in palladium compounds was evaluated, and the expanded uncertainties were calculated. Results indicated that the expanded uncertainties are 0.30%, 0.26%, 0.24%, 0.10% and 0.03% for palladium contents of 59.59% (PdCl2), 47.73% (Pd(OAc)2), 42.46% (Pd(NH3)4Cl2), 17.64% (Pd(NO3)2 solution) and 4.01% (PdSO4 solution) in palladium compounds, respectively. The uncertainty mainly came from the end judgment and calibration process.
View Full Text   View/Add Comment  Download reader
Close