Abstract
醋酸酐生产用废铑催化剂中铑的测定
The Determination of Rhodium in Waste Rhodium Catalyst for Production of Acetic Anhydride
Received:May 25, 2018  
DOI:
中文关键词: 分析化学    密闭消解  分光光度法  电感耦合等离子体发射光谱法(ICP-AES)
英文关键词: analytical chemistry  rhodium  sealed digestion  spectrophotometry  ICP-AES
基金项目:
Author NameAffiliationE-mail
YI Bingzhi Zhejiang Teli Recycling Resources Co. Ltd., Jiaxing 314300, Zhejiang, China gary_yi@126.com 
CHEN Jianfeng Zhejiang Teli Recycling Resources Co. Ltd., Jiaxing 314300, Zhejiang, China  
YAO Tiantian Zhejiang Teli Recycling Resources Co. Ltd., Jiaxing 314300, Zhejiang, China  
ZHANG Yuhui Zhejiang Teli Recycling Resources Co. Ltd., Jiaxing 314300, Zhejiang, China  
JI Zuocheng Zhejiang Teli Recycling Resources Co. Ltd., Jiaxing 314300, Zhejiang, China  
LUO Cheng Zhejiang Teli Recycling Resources Co. Ltd., Jiaxing 314300, Zhejiang, China  
Hits: 2538
Download times: 2172
中文摘要:
      醋酸酐生产用废铑催化剂样品在密闭消解罐中用HNO3-H2O2于160℃经8 h可消解完全,并避免样品损失。考察了样品处理和不同测定方法对测定结果的影响。结果表明,合适的样品量为0.1~0.3 g;消解液经H2SO4-HClO4发烟处理可以去除样品中有机组分的干扰;氯化亚锡分光光度法和ICP-AES法均可准确测定样品中的铑含量,测定相对标准偏差(RSD)均小于2%,加标回收率分别为97.53%~101.65%和98.19%~102.70%,在实际测定中可根据实验室条件选择合适的测定方法。
英文摘要:
      The waste rhodium catalyst for production of acetic anhydride was digested by HNO3-H2O2 in a sealed digestion tank at 160℃, and the sample was completely digested without loss after 8 hours. The effect of sample pretreatment and different methods on the determination results was investigated. The results showed that the optimum sample weight was 0.1~0.3 g. The digestion solution was treated by H2SO4-HClO4 to remove the interference of organic matter. Both stannous chloride spectrophotometry and ICP-AES methods can be used for determining the rhodium content in samples. The relative standard deviation was less than 2%, and their recovery rates were 97.53%~101.65% and 98.19%~102.70%, respectively. In actual applications, an appropriate method can be selected according to the available laboratory conditions.
View Full Text   View/Add Comment  Download reader
Close