Abstract
ICP-AES测定银内胆水杯中有害元素迁移量
Determination of the Migration of Harmful Elements in Silver Tank Cup by ICP-AES
Received:August 06, 2020  
DOI:
中文关键词: 原子发射光谱法(ICP-AES)  有害元素  银内胆水杯    迁移量
英文关键词: ICP-AES  harmful elements  silver tank cup  antimony  migration
基金项目:天津市市场和质量监督管理委员会科技计划项目(编号:2020-W18)
Author NameAffiliation
NA Bo Tianjin Product Quality Inspection Technology Research Institute, Tianjin 300384, China
National Testing Center for Gold and Silver Jewelry Tianjin, Tianjin 300384, China 
HE Ting-ting Tianjin Product Quality Inspection Technology Research Institute, Tianjin 300384, China
National Testing Center for Gold and Silver Jewelry Tianjin, Tianjin 300384, China 
YIN Yi Tianjin Product Quality Inspection Technology Research Institute, Tianjin 300384, China
The National Center of Supervision and Inspection for Quality of Printing and Decoration Product, Tianjin 300384, China 
ZHANG Li-ming Tianjin Product Quality Inspection Technology Research Institute, Tianjin 300384, China
The National Center of Supervision and Inspection for Quality of Printing and Decoration Product, Tianjin 300384, China 
MENG Jian-hua Tianjin Product Quality Inspection Technology Research Institute, Tianjin 300384, China
National Testing Center for Gold and Silver Jewelry Tianjin, Tianjin 300384, China 
CUI Jian-jun Tianjin Product Quality Inspection Technology Research Institute, Tianjin 300384, China
National Testing Center for Gold and Silver Jewelry Tianjin, Tianjin 300384, China 
CAO Wei-yu Tianjin Product Quality Inspection Technology Research Institute, Tianjin 300384, China
National Testing Center for Gold and Silver Jewelry Tianjin, Tianjin 300384, China 
Hits: 939
Download times: 0
中文摘要:
      建立了一种基于电感耦合等离子体原子发射光谱法(ICP-AES)测定银内胆水杯中6种有害元素(砷、镉、铬、镍、铅和锑)迁移量的方法。结果表明,待测元素在标准曲线浓度范围内均呈良好线性关系,加标回收率86.1%~103.3%,相对标准偏差(RSD)均小于5%。在酸性迁移条件下,部分样品可检出一定量的锑元素迁移量,锑迁移量为0.00868~0.0304 mg/dm2。可为银餐具安全评价提供技术支持。
英文摘要:
      A method based on inductively coupled plasma optic emission spectrometry (ICP-OES) for the determination of the migration of six harmful elements (As, Cd, Cr, Ni, Pb and Sb) in the silver tank cup was established. The results showed that the elements to be tested had a good linear relationship in the concentration range of the standard curve, the recovery rate was 86.1%~103.3%, and the relative standard deviate (RSD) was less than 5%. Under the condition of acid migration, a certain amount of antimony migration can be detected in some samples, and the measured antimony migration was 0.00868~0.0304 mg/dm2. The results of this study can provide technical support for the safety evaluation of silver tableware.
View Full Text   View/Add Comment  Download reader