Abstract
草酸银中银的还原重量法测定
Determination of silver content in silver oxalate by reduction gravimetric method
Received:October 15, 2020  
DOI:
中文关键词: 重量法  水合肼还原  草酸银  银含量
英文关键词: gravimetric method  reduction of hydrazine hydrate  silver oxalate  silver content
基金项目:
Author NameAffiliationE-mail
GUAN You-xiang* Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China
State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals, Sino-Platinum Metals Co.Ltd., Kunming 650106, China 
2432028977@qq.com 
LI Kun Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China
State Key Laboratory of Advanced Technologies for Comprehensive Utilization of Platinum Metals, Sino-Platinum Metals Co.Ltd., Kunming 650106, China 
 
LIU Li Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China  
PENG Chun-sheng Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China  
WANG Lin-na Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China  
CAO Peng-bin Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China  
YIN Yao-feng Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China  
YUAN Yuan Yongxing Sino-Platinum Metals Resources Co.Ltd., Chenzhou 423399, Hunan, China  
Hits: 333
Download times: 190
中文摘要:
      采用水合肼还原重量法测定草酸银中银含量,平行测定5次,测定的相对标准偏差小于0.04%,加标回收率99.96%~100.02%,干扰试验显示:0.1 mg的Zn2+、Co2+、Ni2+、Cu2+、Cd2+不干扰银的测定。该法与其他方法相对比,样品取样量大,操作简单,可满足生产现场测定的要求。
英文摘要:
      The silver content in silver oxalate is determined by the hydrazine hydrate reduction gravimetric method, whose determination is performed 5 times in parallel. The relative standard deviation of the determination is less than 0.04%, and the recovery rate of standard addition is 99.96%~100.02%.The interference test showed that 0.1 mg of Zn2+, Co2+, Ni2+, Cu2+, Cd2+ did not interfere with the determination of silver. Compared with other methods, this method has a large sample size and simple operation, which can meet the requirements of production site measurement.
View Full Text   View/Add Comment  Download reader