电感耦合等离子体发射光谱法测定金饰品中铼元素 |
Determination of rhenium in gold adornment by inductively coupled plasma optical emission spectrometry |
Received:May 27, 2021 |
DOI: |
中文关键词: 电感耦合等离子体发射光谱法 金饰品 铼 谱线 |
英文关键词: ICP-OES gold adornment rhenium spectral line |
基金项目: |
Author Name | Affiliation | NA Bo | Tianjin Product Quality Inspection Technology Research Institute,National Testing Center for Gold and Silver JewelryTianjin, Tianjin 300384, China | WANG Wei | Huajin GuojianShenzhen Testing Center of Gold and Silver Jewelry Co.Ltd., Shenzhen 518000, Guangdong China | CUI Jian-jun | Tianjin Product Quality Inspection Technology Research Institute,National Testing Center for Gold and Silver JewelryTianjin, Tianjin 300384, China | LI Peng | Tianjin Product Quality Inspection Technology Research Institute,National Testing Center for Gold and Silver JewelryTianjin, Tianjin 300384, China | MENG Jian-hua | Tianjin Product Quality Inspection Technology Research Institute,National Testing Center for Gold and Silver JewelryTianjin, Tianjin 300384, China | CAO Wei-yu | Tianjin Product Quality Inspection Technology Research Institute,National Testing Center for Gold and Silver JewelryTianjin, Tianjin 300384, China |
|
Hits: 311 |
Download times: 198 |
中文摘要: |
选取346.046 nm作为铼元素的测试谱线,建立了电感耦合等离子体发射光谱法(ICP-OES)测定金饰品中铼元素含量的方法。在0~5.0 μg/mL线性范围内,标准曲线相关系数0.9984,检出限为0.0224 μg/mL。加标量为1.0, 3.0, 5.0 μg/mL时,相对标准偏差(RSD,n =6)分别为0.64%,0.55%,0.78%,平均加标回收率分别为102.7%,97.7%,93.7%。结果表明,该方法适用于金饰品中铼元素的测定。 |
英文摘要: |
An inductively coupled plasma optical emission spectrometry (ICP-OES) method based on the determination of rhenium in gold adornment was established, 346.046 nm was selected as the test spectral line of rhenium. In the linear range of 0~5.0 μg/mL, the correlation coefficient of the standard curve was 0.9984, and the detection limit was 0.0224 μg/mL. Under the three spiked levels of 1.0, 3.0 and 5.0 μg/mL, the relative standard deviations (RSD, n= 6) were 0.64%, 0.55% and 0.78%, the recovery rate were 102.7%, 97.7% and 93.7%, respectively. The results showed that this method was suitable for the determination of rhenium in gold adornment. |
View Full Text
View/Add Comment Download reader |